Interchangeable Range of Ozone Concentration Simulation for Low Cost Reconfigurable Brass Gas Cell
Loading...
Date
Journal Title
Journal ISSN
Volume Title
Publisher
Penerbit UTM Press, Universiti Teknologi Malaysia.
DOI
Abstract
Ultraviolet absorption spectroscopy is reliable for ozone concentration measurement. Concentration range
and optical path length are inversely related based on theoretical calculation and observation of previous
work. However, gas cells for ozone application are typically not expandable. In addition, they incur cost
for custom fabrication. Here we design a reconfigurable brass gas cell that may interchange optical path
length between 5.6 cm and 10.8 cm. Components are available at low cost, easy to joint and ready to use.
Theoretical background and gas cell structure are discussed. Practical transmittance values between e
-0.65
and e
-0.05
are proposed for theoretical calculation of concentration via Beer-Lambert law. The concentration
values are used in SpectralCalc.com gas cell simulation to obtain transmittance. Both approaches yield
comparable result. Simulation result shows concentration range of 5.6 cm optical path length gas cell
(31.82 ppm to 413.67 ppm) is wider than concentration range of 10.8 cm optical path length gas cell (16.50
ppm to 214.49 ppm). Simulation condition is at transmittance from 0.5291 to 0.9522, sampling wavelength
253.65 nm, temperature 300 K and pressure 1 atm. Thus, we strongly recommend short optical path length
gas cell (5.6 cm) for wide range of concentration measurement (31.82 ppm to 413.67 ppm)