A New Ozone Concentration Regulator

Abstract

Laboratory design of an ozone concentration regulator which is build on the theory of continuity equation for gas flow in parallel pipes in conjunctionwith the ozone elimination potentials of an ozone destructor is presented. At an initial oxygen flow rate of 33.33 cm3s-1, ozone concentration was generated and varied between 429.30 parts per million (ppm) to 3826.60 ppm. Similarly at an initial oxygen flow rate of 25 cm3s-1, ozone concentration was generated and varied between 387.30 ppm to 4435.20 ppm. Effect of flow meter when the ozone concentration was set to approximate 1000 ppm were investigated and reported. Fine tuning of the regulator is necessary to ensure concentration stability for long duration experimental work.

Description

Citation

Endorsement

Review

Supplemented By

Referenced By