Investigation of the Effect of Inlet Radius on the Response Time of a Transmission Type Ozone Sensor

dc.contributor.authorDavid, Michael
dc.contributor.authorEn Marcus, Tay Ching
dc.contributor.authorYaacob, Maslina
dc.contributor.authorHussin, Nabihah
dc.contributor.authorSalim, Mohd Rashidi
dc.contributor.authorIbrahim, Mohd Haniff
dc.contributor.authorIdrus, Sevia Mahdaliza
dc.contributor.authorAzmi, Asrul Izam
dc.contributor.authorNgajikin, Nor Hafizah
dc.date.accessioned2021-07-01T09:24:00Z
dc.date.available2021-07-01T09:24:00Z
dc.date.issued2014-12-08
dc.description.abstractThe effect of inlet radius of a transmission type optical gas cell on its response time is reported. Six gas cells of varying lengths, and internal radius of 0.32cm were considered at first and then other internal diameters were also investigated afterwards. The effect of inlet radius is easily discernible at all velocities considered; however it is more pronounced at lower flow rates. At a velocity of 16.79cm/s of ozone gas, and for a target sensing time of ≤ 0.5 seconds; we observed that the inlet radius requirements for gas cells of varying lengths and varying internal diameters is not the same for a specific target sensing speed. The length and the internal radius of a gas cell are proportional to its inlet radiusen_US
dc.description.sponsorshipThe authors would like to thank Universiti Teknologi Malaysia (UTM) for sponsoring this publication under Research University Grant (RUG) Scheme, grant no: 05J60 and 04H35; and Fundamental Research Grant Scheme (FRGS) grant no: 4F317 and 4F565. The Ministry of education (MOE) Malaysia is also acknowledge and the Nigerian Education Trust Fund (ETF) for the financial support giving inform of Tertiary Education Trust Fund (TET-Fund)en_US
dc.identifier.urihttp://repository.futminna.edu.ng:8080/jspui/handle/123456789/5714
dc.language.isoenen_US
dc.publisher2014 IEEE Asia-Pacific Conference on Applied Electromagnetics (APACE)en_US
dc.subjectOptical path length; inlet diameter; internal diameter; response time; flow rateen_US
dc.titleInvestigation of the Effect of Inlet Radius on the Response Time of a Transmission Type Ozone Sensoren_US
dc.typePresentationen_US

Files

Original bundle

Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
Investigation_of_the_Effect_of_Inlet_Radius.pdf
Size:
1.65 MB
Format:
Adobe Portable Document Format
Description:

License bundle

Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
license.txt
Size:
1.71 KB
Format:
Item-specific license agreed upon to submission
Description: