Design and Optimise a GARField MRI Resonator
Loading...
Date
Journal Title
Journal ISSN
Volume Title
Publisher
Advances in Research
DOI
Abstract
The design of a Nuclear Magnetic Resonance (NMR) sensor coil for a GARField NMR system was
examined. The target design has a diameter about 10 − 20 mm and length 5 − 15 mm tuned to
frequency of 22 − 23.4 MHz at 50 Ω total impedance. Nine different sets of coils were built with
different numbers of turns (3, 5, and 7) and different thickness of wire to vary the wire resistance.
The report was to examine based on the design parameters the best resonant circuit for a GARField
MRI system. The acquired tuning characteristics from these resonant circuits were interpreted using
MATLAB scripts and Excel spreadsheet to compare each coil with already existing theory of
resonators. This was achieved by matching each resonant circuit using a match and tuning
capacitor to the required frequency (22-23.4 MHz) and to 50 Ω total impedance at resonance. It was
found that there is no easy method to estimate the inductance of the coil of wire. The result for the
experimental inductance was found to be 0.5 µF and resistance of 0.4 Ω for a medium coil of wire
with 5 numbers of turns, diameter of 0.45 and length of 0.7 mm. The initial attempt to fit the
experimental data to that of the theory failed due to the absence of stray capacitance in the theory.
However, when stray capacitor with value ranging between 30 − 40 pF was incorporated in parallel
with the tank circuit, it was found that both the experiment and theory fit as
expected.
Three coils were tested in the NMR laboratory using a GARField spectrometer to examine the best
Description
Citation
A. Muhammad, M. D. Oladipupo, I. K. Suleiman, & O. L. Jane (2021): Design and Optimise a GARField MRI Resonator. Advances in Research (AIR), 22(2), pp 12-27.