Please use this identifier to cite or link to this item: http://repository.futminna.edu.ng:8080/jspui/handle/123456789/26232
Title: High Energy X-Ray Dosimetry Using (ZnO)0.2(TeO2)0.8 Thin Film based Real-time X-Ray Sensor.
Authors: Ibrahim, Sharifat Olalonpe
Keywords: Keywords: Thin film; X- ray radiation; I-V characteristics; Dosimetry
Issue Date: Apr-2023
Publisher: Journal of Non-Metallic Material Science volume 5(1).
Citation: M.M. Idris, I.O. Olarinoye, M.T. Kolo and S.O. Ibrahim and J.K. Audu. High Energy X-Ray Dosimetry Using (ZnO)0.2(TeO2)0.8 Thin Film based Real-time X-Ray Sensor. Journal of Non-Metallic Material Science vol 5(1). (2023) https://ojs.bilpublishing.com/index.php/nmms
Series/Report no.: NMMS-V5I1-5369;
Abstract: This study reports the dosimetric response of a (ZnO)0.2(TeO2)0.8 thin film sensor irradiated with high-energy X-ray radiation at various doses. The spray pyrolysis method was used for the film deposition on soda-lime glass substrate using zinc acetate dehydrate and tellurium dioxide powder as the starting precursors. The structural and morphological properties of the film were determined. The I-V characteristics measurements were performed during irradiation with a 6 MV X-ray beam from a Linac. The results revealed that the XRD pattern of the AS-deposited thin film is non-crystalline (amorphous) in nature. The FESEM image shows the non-uniform shape of nanoparticles agglomerated separately, and the EDX spectrum shows the presence of Te, Zn, and O in the film. The I-V characteristics measurements indicate that the current density increases linearly with X-ray doses (0-250 cGy) for all applied voltages (1-6 V). The sensitivity of the thin film sensor has been found to be in the range of 0.37-0.94 mA/cm2/Gy. The current-voltage measurement test for fading normalized in percentage to day 0 was found in the order of day 0 > day 15 > day 30 > day 1 > day 2. These results are expected to be beneficial for fabricating cheap and practical X-ray sensors.
URI: http://repository.futminna.edu.ng:8080/jspui/handle/123456789/26232
Appears in Collections:Physics

Files in This Item:
File Description SizeFormat 
High Energy X-Ray Dosimetry Using (ZnO)0.2(TeO2)0.8 Thin Filmbased Real-time X-Ray Sensor.pdf572.61 kBAdobe PDFView/Open


Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.