Please use this identifier to cite or link to this item: http://repository.futminna.edu.ng:8080/jspui/handle/123456789/10345
Full metadata record
DC FieldValueLanguage
dc.contributor.authorJack, Kufre E.-
dc.contributor.authorEtu, Israel A.-
dc.contributor.authorUkanide, Veronica N.-
dc.date.accessioned2021-07-18T01:53:22Z-
dc.date.available2021-07-18T01:53:22Z-
dc.date.issued2016-10-
dc.identifier.citationKufre Esenowo Jack, Israel Agwu Etu & Veronica Nicholas Ukanide (2016). The design of a signal conditioning and acquisition elements of a chopped broadband radiation pyrometer that is capable of measuring temperature between 900oC and 1200oC. Proceedings of the World Congress on Engineering and Computer Science organized by International Association of Engineers, held at, University of California, San Francisco USA, October, 2016, Vol.1, pp. 338-344.en_US
dc.identifier.urihttp://repository.futminna.edu.ng:8080/jspui/handle/123456789/10345-
dc.description.abstractThis paper presents the design of signal conditioning and acquisition elements of a chopped broadband radiation pyrometer. This instrument is capable of measuring temperature between 900oC and 1200oC. This work aims at solving the problem of measuring hot objects with a thermometer. The radiation pyrometer is a non-contact temperature sensor that infers the temperature of an object by detecting its naturally emitted thermal radiation. It collects the visible and infrared energy and focuses it on a detector. The detector used in this device is a thermal sensor. It receives heat energy reflected from a mirror inclined at 45o to the incident signal from the hot object. The design achieved the following: temperature range measured, from 900 to 1200 ; the calibrated instrument is fairly linear with a tolerable nonlinearity of 3.6%, with the sensitivity of . The resolution was quite very small as such it can easily detect the slightest change at its input; the rotating shutter was configured to supply the chopped signal; it operates at a frequency of 50Hz that is lower than the system frequency of 200Hz. The data acquisition system was able to capture data at a periodic time of 0.02 second and below, the system operates within the specified sampling range thus, satisfying Nyquist criteria. The signal so received by the detector is translated to a human readable form and sent to a display.en_US
dc.publisherProceedings of the World Congress on Engineering and Computer Science 2016 Vol I WCECS 2016, October 19-21, 2016, San Francisco, USAen_US
dc.subjectBroadband Radiation Pyrometeren_US
dc.subjectInstrumenten_US
dc.subjectTemperature Sensoren_US
dc.subjectChoppeden_US
dc.subjectDetectoren_US
dc.titleThe Design of a Signal Conditioning and Acquisition Elements of a Chopped Broadband Radiation Pyrometeren_US